Victor Loop Calibrator 05


Victor Loop Calibrator 05

Artikel-Nr.: 5914

Auf Lager

Alter Preis 199,00 €
189,99
Preis inkl. MwSt., zzgl. Versand
Versandgewicht: 550 g


Product Model:V05

Product Name:V 05 Loop Calibrator

Products:Process Calibrator

Detail:

V05  Loop Calibrator

 

Input

Range

Input range

Resolution

Accuracy

Explanation

Voltage

28V

-0.200 ~28.000V

1mV

±0.02% reading ±2mV

Input resistance 2MΩ

Current

20mA

-1.000~22.000mA

0.001mA

±0.02% reading ±4uA

Input resistance 20MΩ

Loop current

20mA

0.000~22.000mA

0.001mA

±0.02% reading ±4uA

Supply 24V loop power

 

Output

Range

Input range

Resolution

Accuracy

Explanation

Current

20mA

0.000~22.000mA

0.001mA

±0.05% Setting value ±4uA

20mA,the max.overload is 1kΩ(Note1)

Analogue transducer

-20mA

0.000~-22.000mA

0.001mA

±0.05% Setting value ±4uA

20mA,the max.overload is 1kΩ(Note2)

Loop power

24V

 

 

±10%

The max.output current is 25mA

 

Power

9V battery (ANSI/NEDA 1640A or IEC 6LR619 alkaline) or AC power adapter

Battery

Approx. 20 hours under the condition of 10mA

Max.allowable voltage

30V

Operation temperture range

0-50°C

Operation humidity range

≤80%RH

Store temperature range

≤-10°C-50°C

Store humidity range

≤90%RH

Size

200×100×40mm

Weight

550g(with holster)

Accessories

Operation manual,test line CF-36(probe with alligator clip)

Options

AC adapter(VCPS),test line CF-31-A(probe clip)

Security

Conforms to IEC 1010


Note:

1.The max.overlaod is 1kΩ at 20mA range when the power is higher than 6.8V,

The max.overlaod is 700Ω at 20mA range when the power is higher than 5.8-6.8V,

2.Power supply range:DC 5-25V

3.Temperature coefficient:±0.005%,range/°C(5°C-18°C,28°C-40°C)

Diese Kategorie durchsuchen: Kalibrator Process